• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER 10X20CM; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER 10X20CM; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Catalog Number 66800900
Device Problem Material Disintegration (1177)
Patient Problem No Patient Involvement (2645)
Event Date 08/31/2020
Event Type  malfunction  
Event Description
It was reported that the silicone of the allevyn is offsetting, there were silicone residues left in the backing paper.No patient was involved.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition, it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaints history review was carried, there have been further complaints reported with this failure mode in the past three years.The devices were intended for use in treatment.The returned dressings were visually and functionally evaluated which confirmed the issue of silicone offsetting.This confirmed a relationship between the event and the device.The wound contact surface of the allevyn gentle border is coated with a gentle silicone adhesive layer that ensures non-traumatic removal at dressing changes.The silicone adhesive will feel sticky when compared to an acrylic adhesive and is specially designed to be very soft and gentle, but can soften further, particularly at higher temperatures.This is an inherent characteristic of all of the silicone adhesives and gives them their soft / gentle properties.It is possible the silicone adhesive on the wound contact layer has problem, with the root cause established as a raw material issue.No further actions are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith + nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN GENTLE BORDER 10X20CM
Type of Device
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10702082
MDR Text Key212050982
Report Number8043484-2020-03242
Device Sequence Number1
Product Code NAD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800900
Device Lot Number201901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2020
Initial Date Manufacturer Received 09/24/2020
Initial Date FDA Received10/19/2020
Supplement Dates Manufacturer Received10/23/2020
Supplement Dates FDA Received11/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-