Model Number N/A
Device Problem Intermittent Loss of Power (4016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).(b)(6).Review of the most recent repair record determined the unit would only operate intermittently and the motor, switch, and reciprocating arm were replaced and resolved the reported issue.Review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It has been reported that the unit was not powering up before surgery.No harm or delay occurred with the patient.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key10707981
MDR Text Key212280062
Report Number0001526350-2020-00879
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number63247388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/20/2020
Initial Date FDA Received10/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-