Model Number 7842 |
Device Problems
Difficult to Insert (1316); Material Integrity Problem (2978); Positioning Problem (3009)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The product has been received for analysis.This report will be updated should further pertinent information be provided from the analysis.
|
|
Event Description
|
It was reported that this right ventricular (rv) lead was an attempted implant due to a lead body damage.After multiple attempts to reposition the lead, the physician could no longer advance a stylet down the lumen of the lead.He then requested a new one.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
The product has been received for analysis.This report will be updated should further pertinent information be provided from the analysis.Upon receipt at our post market quality assurance laboratory, visual inspection noted the helix was retracted and dried blood/tissue was present around the helix.Detailed analysis confirmed that the inner conductor coil is intact.The difficult to position allegation was not confirmed.Analysis found no damage or defect consistent with placement difficulty.The difficult to insert allegation was not confirmed.A stylet insertion test passed indicating no blockage in the lumen.The damage/defective allegation was not confirmed.No visible damage to the lead tip, lead body or helix.
|
|
Event Description
|
It was reported that this right ventricular (rv) lead was an attempted implant due to a lead body damage.After multiple attempts to reposition the lead, the physician could no longer advance a stylet down the lumen of the lead.He then requested a new one.No adverse patient effects were reported.
|
|
Search Alerts/Recalls
|