Catalog Number UNKNOWN
Device Problem Material Split, Cut or Torn (4008)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/18/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the fecal management system's irrigation line tore 1/2 inch and it resulted to leak.And also it was stated the medication was being instilled.
 
Manufacturer Narrative
The reported event was inconclusive as no sample returned for evaluation.A potential root cause for this failure mode could be due to a thin spot on the tube or variation in wall thickness.It was unknown whether the device had met relevant specifications.The product used for the treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number was unknown therefore the device history record could not be reviewed.The product catalog number and the lot number for this device was unknown.Therefore bd was unable to determine the associated labeling to review.Correction: b, d, e, h.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the fecal management system irrigation line tore a half inch which resulted in leakage.And also stated that the medication was being instilled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN FECAL MANAGEMENT (DIGNISHIELD)
Type of Device
DIGNISHIELD
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10708588
MDR Text Key212265430
Report Number1018233-2020-20251
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
PMA/PMN Number
K133251
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 04/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received04/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;