This is being filed to report the steerable guide catheter (sgc) leak.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with grade 4.The steerable guide catheter (sgc) was advanced but could not cross the septum as that area was thicker.Therefore, it was decided to make a new hole.The sgc was removed and during re-flushing of the sgc, it was observed that the hemostatic valve didn¿t hold the water column.The sgc was exchanged with a new sgc.The new sgc was advanced through a new hole without issued.One clip was implanted, reducing mr to 1.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
|
All available information was investigated and the reported leak (loss of fluid column) could not be confirmed/tested via returned device analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar complaints.Based on the information reviewed the loss of fluid column was due to the silicone valve tear.A cause for the silicone valve tear could not be determined.The reported failure to advance appears to be due to patient conditions.There is no indication of a product issue with respect to manufacture, design or labeling.Na.
|