MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN CAN BONE SCREW 6.5MMX25MM; BONE SCREWS AND PINS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN CAN BONE SCREW 6.5MMX25MM; BONE SCREWS AND PINS : SCREWS Back to Search Results
Model Number 1217-25-500
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 07/21/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Empty packing.It was reported that during the surgery of total hip replacement operation, opened the packing (could not find this issue if did not open the packing), noted no product in the packing.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned packaging materials confirms the reported observation.Based on the condition of the inner packaging materials it is difficult to conclude it was distributed in this condition.No error in initial processing is suspected.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: lot pu154657.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Device history review: a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN CAN BONE SCREW 6.5MMX25MM
Type of Device
BONE SCREWS AND PINS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10713056
MDR Text Key212474386
Report Number1818910-2020-22960
Device Sequence Number1
Product Code NDJ
UDI-Device Identifier10603295009900
UDI-Public10603295009900
Combination Product (y/n)N
PMA/PMN Number
K983014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 07/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1217-25-500
Device Catalogue Number121725500
Device Lot NumberPU154657
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/05/2020
Initial Date Manufacturer Received 08/05/2020
Initial Date FDA Received10/21/2020
Supplement Dates Manufacturer Received10/22/2020
Supplement Dates FDA Received10/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-