This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device's spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
It was reported that this implantable cardioverter defibrillator (icd) device, and non-boston scientific leads exhibited one episode of high, out of range pacing impedance greater than 2,000 ohms.There were no other out of range measurements, and no noise present.A technical service (t/s) consultant, reviewed the device data and suggested programming options, and lead integrity testing.It was also noted, that the analysis showed the patient to be exhibiting atrial fibrillation.T/s recommended following patient closely over a home monitor equipment.The device remains in service.No adverse patient effects were reported.
|