Medtronic received a report that an abnormality was noted on the behavior of the catheter and guidewire under fluoroscopy while the catheter was being delivered to the target lesion.After being retrieved, it seemed the guidewire came out from the proximal part of the tip of the catheter.After the inside of the catheter was flushed, it was confirmed that the catheter was perforated in the proximal part of the distal tip.There hadn't been any resistance when passing the guidewire through, and there was no damage observed to the guidewire.Replacement products were used to complete the procedure.It was indicated that all devices were prepared as per the instructions for use (ifu), and it was reported there was no patient involvement associated with the event.Ancillary devices include a shuttle sheath 5f, tenrou10 guidewire.
|