• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Intermittent Continuity (1121); Gas Output Problem (1266); Failure to Deliver (2338)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/30/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that the device had a ventilator failure during use.There was no patient injury reported.
 
Manufacturer Narrative
The responsible dräger service engineer could confirm the reported issue during on-site checking and trace it back to the ventilator motor.The entire motor assembly was replaced, consequently.Also the evaluation of the log file confirms an issue with the ventilator (motor) on the date of event.It was found that the device forced a shutdown of automatic ventilation due to a detected wrong motor position.The motor speed is being monitored continuously; speed fluctuations caused e.G.By an abraded collector disc will result in a deviation between measured and expected piston position.To prevent from damages, the system is designed to shut down automatic ventilation and to alert the user to this condition by means of a corresponding alarm.Manual ventilation and the monitoring functions remain available to the full extent.Dräger finally concludes that the device behaved as specified upon the malfunction of a single component; no patient consequences have been reported.The repair exchange of the motor unit has fully solved the problem.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
It was reported that the device had a ventilator failure during use.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key10716794
MDR Text Key212481547
Report Number9611500-2020-00369
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/05/2020
Initial Date FDA Received10/22/2020
Supplement Dates Manufacturer Received10/05/2020
Supplement Dates FDA Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-