Model Number 131F7
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/01/2020
Event Type  malfunction  
Manufacturer Narrative
One 131f7 with a monoject limited volume syringe was returned for examination.The reported event of "detached" was confirmed.As received, the sheath was not returned.The catheter balloon was found to be torn from the proximal and distal bonding sites.The central area of balloon latex was missing.Bonding indication and residual latex were observed from the distal and proximal bonding sites.All through lumens were patent without any leakage or occlusion.No visible damage or inconsistency was observed from the catheter body.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is standard practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.When there is separation of the balloon or fragments from the pulmonary artery catheter, the retained fragment can embolize to the lungs.Due to the large surface area of the pulmonary vasculature, this is generally well tolerated, but can lead to complications such as infection or small infarction.Pulmonary complications may result from improper inflation technique.To avoid damage to the pulmonary artery and possible balloon rupture, the balloon should not be inflated above the recommended volume.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon detached during a case.The customer indicated that the catheter came out to the sheath with the balloon sheared off.They could not find it so it was assumed that the balloon remained in the body.The patient is stable and was told of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ THERMODILUTION CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492502904
MDR Report Key10718230
MDR Text Key213186161
Report Number2015691-2020-14068
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K810124
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/10/2022
Device Model Number131F7
Device Catalogue Number131F7P
Device Lot Number62937457
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
Patient Weight132
-
-