• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P RSP; RSP BASEPLATE, 30MM, W/P2 COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P RSP; RSP BASEPLATE, 30MM, W/P2 COATING Back to Search Results
Model Number 508-32-204
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 09/25/2020
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Third revision surgery - the patient broke the center screw on his original implant.
 
Manufacturer Narrative
The reason for this revision surgery was reported as broken center screw.The previous surgery and the surgery detailed in this event occurred 8 months and 1 week apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to broken center screw.There were no other findings during this evaluation that indicate the reported devices were defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, prolonged overhead activities, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
RSP BASEPLATE, 30MM, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin,tx 78758-5445
MDR Report Key10726616
MDR Text Key212728659
Report Number1644408-2020-00941
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912144391
UDI-Public(01)00888912144391
Combination Product (y/n)N
PMA/PMN Number
K112069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number508-32-204
Device Catalogue Number508-32-204
Device Lot Number769P1730
Was Device Available for Evaluation? No
Date Manufacturer Received11/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
506-03-126 LOT 834C1632; 506-03-130 LOT 835C1610; 506-03-126 LOT 834C1632; 506-03-130 LOT 835C1610
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age77 YR
-
-