It was reported that this left ventricular (lv) lead exhibited high capture thresholds, failure to capture, noise, oversensing, pacing inhibition and pacing impedance high due to lead fracture.The patient also experienced pain due to this issue.Additional information indicates that during removal of this lead, the distal end of the lead separated, leaving the end of the lead in the left ventricle coronary sinus branch.The left ventricular (lv) lead ring electrode separated from the broken end of the lead, traveled up through the pulmonary artery and is believed to be stuck against the pulmonic valve.A team of surgical physicians decided not to retrieve the electrode at this time due to the risk being higher than the benefit for the patient.This lead was successfully replaced.No additional adverse patient effects were reported.The complaint device was not returned by the customer; therefore, no product analysis could be performed.
|