• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.Linked to mfg report numbers: 3004608878-2020-00629 and 3004608878-2020-00628.
 
Event Description
This is 2 of 3 reports.It was reported that the locking knob of the a1059 mayfield modified skull clamp was broken.There was no known patient injury or surgery delay.
 
Manufacturer Narrative
Unique device identification(udi): (b)(4).Mayfield skull clamp was returned for evaluation.Device history record (dhr) - the dhr shows no abnormalities related to the reported failure.The reported complaint was confirmed from the evaluation of the returned product.The lock has rotational and lateral movement and a residue buildup is present.Unit need new components added to replace worn internal parts.The observed condition is likely caused by wear and tear.The definite root cause cannot be reliably determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10730919
MDR Text Key213043529
Report Number3004608878-2020-00627
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Initial Date Manufacturer Received 09/30/2020
Initial Date FDA Received10/23/2020
Supplement Dates Manufacturer Received11/04/2020
Supplement Dates FDA Received11/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-