Catalog Number A1059
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
An inventory control analyst reported that the index knob of the a1059 mayfield modified skull clamp was very difficult to turn to the locked position.This was discovered during a routine inspection.
 
Manufacturer Narrative
Udi: (b)(4).Device history record (dhr) - dhr review shows no abnormalities related to the reported failure.The reported complaint was confirmed from the evaluation of the returned mayfield skull clamp.The lock having movement and a residue buildup is present.Parts of the device are worn and need to be replaced.The observed condition is likely caused by wear and tear of the device.The definite root cause cannot be reliably determined.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10731025
MDR Text Key213043139
Report Number3004608878-2020-00633
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Manufacturer Received 10/05/2020
Initial Date FDA Received10/23/2020
Supplement Dates Manufacturer Received11/17/2020
Supplement Dates FDA Received12/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-