• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NAKANISHI INC. NSK; HANDPIECE, ROTARY BONE CUTTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NAKANISHI INC. NSK; HANDPIECE, ROTARY BONE CUTTING Back to Search Results
Model Number SGS-E2S
Device Problem Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Date 09/04/2020
Event Type  Injury  
Manufacturer Narrative
The distributor is attempting to obtain further information on the event, including information about the patient.Upon receiving the device involved in the mdr event from the distributor, nakanishi conducted a failure analysis of the returned device (b)(4).These activities are described in more detail below.Methodology used: nakanishi examined the device history record and the repair history for the subject sgs-e2s device [serial number: (b)(4)].There were no problems observed during the manufacturing or testing noted in the dhr.There were also no repair history records since the device was shipped.Nakanishi set a test bur in the handpiece, connected the handpiece to the motor, and tried to rotate the motor.However, the handpiece was locked and the motor did not rotate at all.Therefore, nakanishi was not able to conduct temperature testing of the device.Identification of the specific failure mode(s), and/or mechanism(s) of the associated device components was conducted as follows: nakanishi disassembled the handpiece and performed a visual inspection of the internal parts.Nakanishi observed the following phenomena: the gear shaft did not rotate at all.There was debris on the bearing.The internal parts were corroded.Nakanishi took photographs of all of the disassembled parts and kept them in the investigation report # (b)(4).Conclusions reached based on the investigation and analysis results: nakanishi could not replicate the temperature increase from the event, but based on the findings in the visual inspection, as well as many years of experience, nakanishi identified the cause of the handpiece overheating was abnormal resistance during rotation due to the soiled bearing caused by ingress of foreign materials into the handpiece.A lack of maintenance causes the accumulation of debris on the inside parts, which causes debris ingress into the bearing during rotation.This contributes to the handpiece overheating.In order to prevent a recurrence of the handpiece overheating, nakanishi took the following actions: nakanishi reviewed the operation manual, reconfirmed clarity, and understandability of the instructions.Nakanishi reported the above evaluation results to nsk america and directed nsk america to remind the user of the importance of maintenance, as instructed in the operation manual.
 
Event Description
On october 3, 2020, nakanishi became aware of a handpiece overheating through a complaint input into the complaint database by a distributor (nsk america).Details are as follows: the event occurred on (b)(6) 2020.A dentist was performing a dental procedure on a patient using the sgs-e2s handpiece (serial no.: (b)(4)).During the procedure, the handpiece overheated and burned the patient's inner cheek.
 
Manufacturer Narrative
On (b)(6) 2020, nam updated nakanishi about the event.Despite the following attempts, nam received no response from the dentist.On (b)(6) 2020, nam made a phone call to the dental office for detailed information about the event.Nam left a voicemail message requesting the dental office to contact nam.On (b)(6) 2020, nam called the office again and left another message on the voicemail.On (b)(6) 2020, nam sent an e-mail to the dentist requesting the details.On (b)(6) 2020, nam sent another e-mail to the dentist.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NSK
Type of Device
HANDPIECE, ROTARY BONE CUTTING
Manufacturer (Section D)
NAKANISHI INC.
700 shimohinata
kanuma-shi, tochigi-ken 322-8 666
JA  322-8666
MDR Report Key10732463
MDR Text Key216112379
Report Number9611253-2020-00040
Device Sequence Number1
Product Code KMW
Combination Product (y/n)N
PMA/PMN Number
K171155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Remedial Action Recall
Type of Report Initial,Followup
Report Date 01/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSGS-E2S
Device Catalogue NumberH266
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2020
Initial Date Manufacturer Received 10/03/2020
Initial Date FDA Received10/26/2020
Supplement Dates Manufacturer Received12/19/2020
Supplement Dates FDA Received01/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number9611253-060818-001-R
Patient Sequence Number1
Patient Outcome(s) Other;
-
-