During the procedure, a clinically significant delay was noted.The ablation catheter was advanced into the left atrium (la) and ablation was performed without issue.The catheter was withdrawn from the la and a sl0 sheath was advanced into the right atrium (ra).The ablation catheter was then removed and reinserted through the sl0 sheath into la and no contact force values were observed.The catheter was unplugged, plugged back in and the reset button on the tactisys was pressed and a long ¿boop¿ sound was heard.Another catheter was used to complete the procedure with no consequences to the patient.The physician stated that troubleshooting caused a significant delay (approximately 12 minutes).
|
Additional information: d10, g4, h2, h3, h6.One bi-directional, curve d-f, tacticath sensor enabled contact force ablation catheter was received for evaluation.Optical fiber 3 no longer met specifications for optical signal properties and no contact force was displayed when the returned device was connected to a tactisys unit.Further investigation revealed optical fiber 3 was fractured at the distal end of the handle under the strain relief, consistent with the optical signal issue observed and the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the optical fiber fracture is consistent with damage during use.
|