• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG UNKNOWN VENOVO STENT; VENOUS STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG UNKNOWN VENOVO STENT; VENOUS STENT Back to Search Results
Catalog Number UNK VENOVO STENT
Device Problems Material Deformation (2976); Insufficient Information (3190)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2020
Event Type  malfunction  
Manufacturer Narrative
A review was not performed because the lot number was not known.It was therefore not known whether the reported issue was potentially related to manufacturing.The stent delivery system was not available for evaluation; one image was provided demonstrating a released stent inside a vessel with irregular strut pattern and with blood flow which led to a confirmed result for deformation.Based on the information available a definite root cause for the reported event could not be determined.
 
Event Description
It was reported that during treatment, the stent allegedly failed to work as intended.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN VENOVO STENT
Type of Device
VENOUS STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key10734769
MDR Text Key213046024
Report Number9681442-2020-00230
Device Sequence Number1
Product Code QAN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK VENOVO STENT
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/19/2020
Initial Date FDA Received10/26/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-