DEPUY ORTHOPAEDICS INC US PFC*CALIBRATED PAT CUT GDE; KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS Back to Search Results
Model Number 86-5034
Device Problem Structural Problem (2506)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 10/07/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: examination of the returned device could not confirm the reported condition.The investigation found no evidence at this time of product malfunction or product error and the need for corrective action was not established.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Event Description
It was reported that the lobster claw won't lock anymore making it difficult to use.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key10736506
MDR Text Key213080754
Report Number1818910-2020-23333
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295216599
UDI-Public10603295216599
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number86-5034
Device Catalogue Number865034
Device Lot NumberSO2029025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/07/2020
Initial Date FDA Received10/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-