Migration is a known cause of loss of therapy.The patient decided to have the stimulators explanted rather than have a revision to correct the migration.When explanting the stimulator, the physician noticed lead electrode breakage and migration, possibly intrathoracically.The lead was removed during the procedure, with no further issues reported.The remainder of the stimulator was fully retrieved.Based on this information, the fractured stimulator was confirmed/replicated; there is no evidence that the product did not meet specifications.The stimulator is used for the treatment of pain.The cause of the fractured stimulator is unknown/no problem found.
|
The patient was implanted with two stimulators (fr4-a000 and fr4-a000).The patient received good stimulation after the implant procedure.However, on (b)(6) 2019, the patient was implanted with a new lead.The reason why the patient needed an additional lead implanted is unknown.An x-ray confirmed migration.
|