Model Number CD3371-40QC
Device Problem Migration (4003)
Patient Problems Fatigue (1849); Twiddlers Syndrome (2114)
Event Date 10/09/2020
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2017865-2020-16191.It was reported that the patient presented to the emergency room complaining of weakness.Interrogation of the implantable cardioverter defibrillator (icd) revealed that the right ventricular (rv) lead was unable to pace.An x-ray revealed that the rv lead had dislodged into the superior vena cava, and that the icd was pointing in the opposite direction as it did when first implanted, indicating twiddler's syndrome.The device was reprogrammed and the patient was monitored until the rv lead revision occurred.During the procedure, the lead was attempted to be repositioned but the helix was difficult to retract and extend, so the lead was cut, explanted, and replaced.The patient was in stable condition throughout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key10741709
MDR Text Key213245791
Report Number2017865-2020-16194
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberCD3371-40QC
Device Lot NumberP000098842
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM
-
-