• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US XIA 3 TITANIUM POLYAXIAL SCREW 8.5 X 50MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US XIA 3 TITANIUM POLYAXIAL SCREW 8.5 X 50MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 482318550
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2020
Event Type  malfunction  
Manufacturer Narrative
Device remains implanted in patient.
 
Event Description
It was reported that the tulip of a xia 3 titanium polyaxial screw at l4 disengaged less than a month post-operatively.Revision surgery is not scheduled at this time.
 
Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device and complaint history records review could not be performed as a valid lot code was not provided and could not be obtained.No further investigation can be performed due to insufficient information provided.If more information is received and/or devices are returned for evaluation this investigation will be reopened and updated.
 
Event Description
It was reported that the tulip of a xia 3 titanium polyaxial screw at l4 disengaged less than a month post-operatively.Revision surgery is not scheduled at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM POLYAXIAL SCREW 8.5 X 50MM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key10742591
MDR Text Key213478885
Report Number3005525032-2020-00042
Device Sequence Number1
Product Code NKB
UDI-Device Identifier04546540598608
UDI-Public04546540598608
Combination Product (y/n)N
PMA/PMN Number
K142381
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number482318550
Device Catalogue Number482318550
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/29/2020
Initial Date FDA Received10/27/2020
Supplement Dates Manufacturer Received11/30/2020
Supplement Dates FDA Received12/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-