Model Number 71992-01 |
Device Problem
High Readings (2459)
|
Patient Problems
Hypoglycemia (1912); Seizures (2063); Malaise (2359); Loss of consciousness (2418); Sweating (2444); Shaking/Tremors (2515)
|
Event Date 10/14/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A high readings issue was reported with the freestyle libre 2 sensor.Customer reported receiving a higher sensor result when comparing to a reading obtained on the built-in reader.Customer experienced malaise, sweating, shaking, seizure, and a loss of consciousness and required treatment of fruit juice by a third-party.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.
|
|
Event Description
|
A high readings issue was reported with the freestyle libre 2 sensor.Customer reported receiving a higher sensor result when comparing to a reading obtained on the built-in reader.Customer experienced malaise, sweating, shaking, seizure, and a loss of consciousness and required treatment of fruit juice by a third-party.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|