One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.The device met specifications for temperature testing and optical signal properties.Electrical insulation between conductor wires 1-4 and the thermocouple wires was also tested while the catheter was deflected and un-deflected.Short circuits were detected between electrode 1 and electrode 2, electrode 1 and electrode 3, electrode 2 and electrode 3, and electrode 2 and electrode 4.Electrode rings 2-4 were subsequently dissected for further inspection.No breaks in the shaft were noted under electrode rings 2 and 3 but there was a break in the shaft at electrode 4.The 19-pin redel connector was disassembled.Evidence of dried saline was noted within the connector, consistent with the observed short circuits.The catheter shaft was cut proximal to the deflectable section and the tip was removed from the distal shaft.The pi irrigation tubing was found to remain attached to the metal irrigation tubing at the deformable body.The pi irrigation tubing was inspected, and it was noted that the adhesive went all the way around the inside of the tubing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the tear in the shaft material at electrode ring 4 and the resulting short circuits remains unknown.
|