• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22306D
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Burn(s) (1757); Internal Organ Perforation (1987)
Event Type  Injury  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patientâ¿¿s outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection of a bladder tumor (turbt) procedure at an unknown date, the patient sustained a perforation in the intestinal tract as well as a pinpoint burn to bladder and peritoneum.Reportedly, the patient's urethra was very narrow and the surgeon therefore used only the inner sheath without the outer sheath.The patient is currently recovering.No further information was provided.
 
Manufacturer Narrative
The suspect medical device was not returned to the manufacturer for evaluation/investigation since it was reportedly discarded by the user facility.Therefore, the cause of the reported event could not be conclusively determined.Since there was no report about any product malfunction, an anatomical / procedural complication during operation is assumed.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key10762297
MDR Text Key213878364
Report Number9610773-2020-00253
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051689
UDI-Public14042761051689
Combination Product (y/n)N
PMA/PMN Number
K100275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberWA22306D
Device Catalogue NumberWA22306D
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/28/2020
Initial Date FDA Received10/30/2020
Supplement Dates Manufacturer Received05/06/2021
Supplement Dates FDA Received05/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNSPECIFIED VIO HF GENERATOR; UNSPECIFIED VIO HF GENERATOR
Patient Outcome(s) Required Intervention;
-
-