• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD Back to Search Results
Model Number MN10450-50A
Device Problem Migration (4003)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/15/2020
Event Type  Injury  
Manufacturer Narrative
Date of the event is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer reference number: 1627487-2020-33093, 1627487-2020-33094, 1627487-2020-33096.It was reported the patient fell off their bed and 2 of their leads migrated.Reprogramming was unable to cover the pain areas.Surgical intervention took place wherein the 2 leads were explanted and replaced to address the issue.It is unknown which 2 leads had migrated, as such all leads are being reported.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.  based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information indicates the patient is having effective relief again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG SLIM TIP LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key10762498
MDR Text Key213868899
Report Number1627487-2020-33095
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067027153
UDI-Public05415067027153
Combination Product (y/n)N
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/16/2021
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number7219449
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MODEL # 10450-50A - LEAD (X3); MODEL # 10450-50A - LEAD (X3)
Patient Outcome(s) Other;
Patient Weight82
-
-