One used list# 423280412, safeset¿ 24 inch arterial pressure tubing, safeset¿ reservoir and blood sampling port, possible lots 4751943 or 4853674 was received for evaluation.The reported complaint of blood sampling ports came apart was confirmed on the returned set.During visual inspection, the safeset reservoir was found separated from the one way stopcock.The one way stop cock was microscopically examined under a uv light, and insufficient adhesive was observed.The probable cause of the safeset reservoir separation had occurred due to insufficient adhesive applied during assembly process.A device history review for possible lots 4751943, 4853674 and relevant commodities were reviewed, and no non-conformances were found that would have contributed to the reported complaint.The lot number of the device that was in use is unknown.The customer identified two possible lot numbers (plots).The possible lot numbers are 4751943 (expiry date: 03/01/2023, mfr date 03/01/2020), and 4853674 (expiry date 05/01/2023, mfr date 05/01/2020).
|
The event involved a safeset¿ 24 inch arterial pressure tubing, safeset¿ reservoir, and blood sampling port where the blood sampling port came apart during set up, prior to patient care.There was no patient involvement, and no adverse event.
|