• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Failure to Cut (2587)
Patient Problem No Patient Involvement (2645)
Event Date 10/05/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.
 
Event Description
It was reported that during testing the device was cutting grafts in half.There was no harm to the patient, no delay.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor was erratic, control bar not in the correct position, calibration out at all readings, width plate, control bar, and head damaged and the motor, width plate, control bar and head were replaced and device calibrated and resolved the reported issue.It was noted that the device has not been returned for annual pm since it was manufactured.Device is used for treatment.The event cannot be confirmed.
 
Event Description
There is no additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key10762739
MDR Text Key213876899
Report Number0001526350-2020-00910
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number30343400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Was the Report Sent to FDA? No
Date Manufacturer Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-