Model Number N/A |
Device Problem
Power Problem (3010)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 10/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Telephone no: (b)(6).Once the investigation is complete, a follow up/final report will be submitted.
|
|
Event Description
|
It was reported that before surgery the device was not maintaining consistent speed.There was no harm to a patient or operator and no delay in a procedure.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
There is no additional information.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the motor speed was low; however, the repair was not completed because four fillister screws were corroded and stuck inside.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|