• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Device Displays Incorrect Message (2591)
Patient Problem Seizures (2063)
Event Date 10/15/2020
Event Type  Injury  
Manufacturer Narrative
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported receiving a "replace sensor" message after 3 days of wearing the adc freestyle libre 2 sensor.Customer further reported he experienced seizures, however no treatment was rendered.Customer was unsure if the seizures was related to sensor issue.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
Sensor (b)(6) has been returned and investigated.Visual inspection was performed and no issues were observed.Data was extracted from returned sensor using approved software.Sensor was found to be in state 5 (indicating normal termination).The sensor plug was properly seated in the mount.The sensor plug was removed and the plug assembly was inspected, no issues were observed.However, visual inspection was performed on the returned sensor tail.No blood watermark was observed on the tail, indicating an insertion failure.An insertion failure is, due to the sensor not being properly inserted.No malfunction or product deficiency was identified.Section d3: (email), and section g1: were updated to reflect current contact information.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported, receiving a "replace sensor" message after (b)(6) days of wearing the adc freestyle libre 2 sensor.Customer further reported, he experienced seizures.However, no treatment was rendered.Customer was unsure if the seizures was related to sensor issue.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10763669
MDR Text Key213914086
Report Number2954323-2020-09304
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 03/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2022
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Date Manufacturer Received02/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age47 YR
Patient Weight126
-
-