Model Number 9-PDA2-06-06 |
Device Problems
Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/02/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2020, a 6x6mm amplatzer duct occluder ii (adoii) was selected for implant.During the procedure, prior to the deploy of the adoii, the non-abbott sheath got caught between two mitraclips that were implanted approximately 2 months ago.The occluder was intended for implant between the 2 clips; further described as a2p2.The non-abbott sheath was cut at the groin and the sheath and occluder remain in the anatomy of the patient, still caught within the clips.The physician reports no plans to remove either products.A clinically significant prolonged procedure delay was reported.The patient was reported to be in stable condition.
|
|
Manufacturer Narrative
|
An event of a sheath being used to implant a amplatzer duct occluder 2 in between two mitraclips and the sheath and occluder being left in the patient was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the instructions for use, arten600038220 revision a "the amplatzer¿ duct occluder ii is a percutaneous transcatheter occlusion device intended for the non-surgical closure of patent ductus arteriosus".
|
|
Search Alerts/Recalls
|