• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER II; OCCLUDER, PATENT DUCTUS, ARTERIOSUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER II; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDA2-06-06
Device Problems Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2020
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020, a 6x6mm amplatzer duct occluder ii (adoii) was selected for implant.During the procedure, prior to the deploy of the adoii, the non-abbott sheath got caught between two mitraclips that were implanted approximately 2 months ago.The occluder was intended for implant between the 2 clips; further described as a2p2.The non-abbott sheath was cut at the groin and the sheath and occluder remain in the anatomy of the patient, still caught within the clips.The physician reports no plans to remove either products.A clinically significant prolonged procedure delay was reported.The patient was reported to be in stable condition.
 
Manufacturer Narrative
An event of a sheath being used to implant a amplatzer duct occluder 2 in between two mitraclips and the sheath and occluder being left in the patient was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the instructions for use, arten600038220 revision a "the amplatzer¿ duct occluder ii is a percutaneous transcatheter occlusion device intended for the non-surgical closure of patent ductus arteriosus".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER II
Type of Device
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10763773
MDR Text Key213918756
Report Number2135147-2020-00457
Device Sequence Number1
Product Code MAE
UDI-Device Identifier00811806011264
UDI-Public00811806011264
Combination Product (y/n)N
PMA/PMN Number
P020024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-PDA2-06-06
Device Catalogue Number9-PDA2-06-06
Device Lot Number7578528
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age88 YR
Patient Weight73
-
-