Results: the jet7 was kinked approximately 3.5 and 80.0 cm from the hub.The marker band was damaged at the distal tip of the jet7.The entire length of the returned jet7 was intact.Conclusions: evaluation of the returned jet7 was unable to confirm the reported distal fracture.The entire length of the returned jet7 was intact.Further evaluation revealed the distal markerband was damaged.This damage may have been due to handling during use.Additionally, there were kinks on the proximal and mid-shaft of the device.This damage was likely incidental to the reported complaint and may have occurred during packaging for return to penumbra.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7), a non-penumbra microcatheter, a neuron max 6f 088 long sheath (neuron max), a guidewire and a penumbra system ace 68 reperfusion catheter (ace68).During the procedure, the physician advanced the jet7 and microcatheter over the guidewire through the neuron max and into the target vessel to aspirate the clot with no reported resistance.During aspiration while retracting the jet7, the catheter fractured distally.Therefore, the physician removed the jet7 and the fractured tip remained in the patient.Upon removal, it was noticed that the proximal end of the jet7 was kinked.The procedure was completed using an ace68, the same neuron max and microcatheter.However, the fractured distal tip of the jet7 was displaced into the anterior cerebral artery (aca) and was left in the patient¿s body as it was too distal to be retrieved.There was no report of an adverse effect to the patient.After the procedure, the jet7 fractured at the mid-shaft of while it was being packed into the box.
|