As part of the investigation, olympus followed up with the user facility to obtain additional information/clarification on the reported event but with no results.The original equipment manufacturer (oem), performed a device history record review and no abnormalities were noted.The actual product was manufactured on january 10, 2019.A review of the service records show the scope was purchased on (b)(6) 2019 and was last serviced via repair on (b)(6) 2020.No device was returned to the oem and insufficient information was provided; therefore, the oem was unable to investigate the cause leading to the occurrence of this incident.The instructions for user provides the following information: chapter 3 preparation and inspection before each case, prepare and inspect this instrument as instructed below.Inspect other equipment used with this instrument as instructed in their respective instruction manuals.Should the slightest irregularity be suspected, do not use the instrument, see chapter 9, ¿troubleshooting¿.If the irregularity is still suspected after consulting chapter 9,contact olympus.Damage or irregularity may compromise patient or user safety and may result in more-severe equipment damage.This instrument was not disinfected or sterilized before shipment.Before using this instrument for the first time,reprocess it according to the instructions in chapter 5,¿reprocessing: general policy¿ through chapter 7,¿cleaning, disinfection and sterilization procedures.
|