Related manufacturer report number: 2184149-2020-00181, 3008452825-2020-00588.During the procedure, the communication between ensite and ampere and tactisys was unstable, and when the catheter was connected, communication with ampere and tactisys was disconnected.When the tactisys was rebooted, communication was sometimes possible and sometimes not, and when the tacticath was connected, communication was stopped again.Since the same phenomenon occurred multiple times, the tactisys was repeatedly rebooted and eventually was able to communicate.However, because the study was stopped for more than 30 minutes, propofol was additionally administered to the patient.The contact force value was about 10 g before the rf delivery, but it increased rapidly to about 30 g immediately after the rf delivery.The procedure was completed without replacing the devices and there were no adverse consequences to the patient.
|