• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC POWEREASE INSTRUMENTS; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC POWEREASE INSTRUMENTS; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number 2342305M
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/12/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from multiple sources (manufacturer representative, healthcare provider, user facility) regarding a patient had previous fusion (tlif).It was reported that the bone was hard and the driver sheared off when doctor tried to upsize the screw.The product will be returned.There were no patient symptoms or complications as a result of this event.
 
Manufacturer Narrative
H6: product analysis: visual and optical examination confirmed approximately 3mm of instrument tip has been broken off, consistent with interface during usage.Optical examination revealed a fairly flat surface with circular material flow.The remaining tip has been twisted.This type of damage is consistent with torsional overload.Additional information: d9, g3, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWEREASE INSTRUMENTS
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10768376
MDR Text Key214079102
Report Number1030489-2020-01529
Device Sequence Number1
Product Code HBE
UDI-Device Identifier00613994905697
UDI-Public00613994905697
Combination Product (y/n)N
PMA/PMN Number
K111520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2342305M
Device Catalogue Number2342305M
Device Lot NumberCT16L039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received11/02/2020
Supplement Dates Manufacturer Received02/04/2021
Supplement Dates FDA Received02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age61 YR
-
-