Model Number N/A |
Device Problems
Difficult to Insert (1316); Off-Label Use (1494)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/07/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Associated products : item#:42542007501; tibia fixed non-porous left size f; lot#: 64495998.Foreign : (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that total knee arthroplasty was performed with persona.The surgeon could not insert persona mc surface to tibia base plate and he disengaged the surface to clean space between the surface and the tibia base plate by saline.After that, he tried to insert the surface again but he could not.Subsequently, persona uc surface was used to complete the procedure.
|
|
Manufacturer Narrative
|
D11: associated products : item: 42542007501; tibia fixed cemented left size f; lot: 64495998.This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual examination of the returned product identified the dovetail feature exhibits damage (flared out).Device history record (dhr) was reviewed and no discrepancies were found.Per persona revision knee system ifu, it is stated do not use the persona revision femoral components or tibial plates with the persona cruciate retaining (cr), medial congruent (mc) or ultra congruent (uc) articular surfaces.Do not use the persona revision cck articular surfaces with the persona ps or cr femoral components.They are not designed to be compatible.Hence the root cause is attributed to off label use.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2021 - 00400.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|