• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED 7.0; OXYGENATOR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HLS SET ADVANCED 7.0; OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050USA #HLS SET ADVANCED 7.0
Device Problem Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
A follow-up emdr will be submitted when additional information becomes available.
 
Event Description
On (b)(6) we had to exchange a new circuit for our pt.The delta p and p-int were rising so dr gottner wanted replaced the circuit.He also wanted it to be evaluated by the vendor.My assumption he was worried about any malfunction in the system or if you could identify a clot.Complaint id: (b)(4).
 
Manufacturer Narrative
The review of the non-conformities till 2020-11-13 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences can be excluded.The patient was infected with covid-19.Covid-19 diseases can be associated with intravascular coagulation activation, microcirculation disorders and increased risk of thromboembolism despite good systemic anticoagulation.A similar case was already investigated under complaint#(b)(4) on 2020-06-05: as stated in the investigation report of (b)(4) clots could be found inside the oxygenator.The most probable root cause for the increased pressure drop are clots in the oxygenator leading to a blockage and thus an extension of the diffusion path lowering the oxygenating performance.With reference to the risk assessment (hls set advanced 5.0 / hls set advanced 7.0, dms # 1468452, v24) and in consultation with the manager medical affairs the following events can contribute to clotting in the circuit: -de-airing luer lock connection too loose ; -air remains in or enters the circuit; -hemostasis ; -air or blood remains in luer lock access port; -too low anticoagulation; -too low at level, effect of heparin is too limited ; -protamine sulfate enters the hls set; -administration of substitution of congealable substance such as plateles; -(consumption) coagulopathy; -thrombozytopenia.Based on this the reported failure could be confirmed.The occurrence rate was calculated for the reported failure and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLS SET ADVANCED 7.0
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key10770661
MDR Text Key214118077
Report Number8010762-2020-00369
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
PMA/PMN Number
K112360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBEQ-HLS 7050USA #HLS SET ADVANCED 7.0
Device Catalogue Number701052794
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/26/2020
Initial Date FDA Received11/02/2020
Supplement Dates Manufacturer Received10/29/2020
Supplement Dates FDA Received11/25/2020
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-