• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Injury (2348)
Event Type  Injury  
Manufacturer Narrative
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported the perforator failed to disengage, tearing the dura.No additional information has been provided.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: d4: udi: (b)(4).Dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - complaint not confirmed, the returned unit was found to work as intended, and met all acceptance criteria.Possible root causes for drill fails to disengage are: product cannot withstand multiple sterilizations and uses, impact introduction of drill to skull causes scoring on the pin, inadequate spring, drill used for multiple holes and chatter/deformation of pin / slot interface occurs, drill set incorrectly, or incorrect specification for surface finish for inner drill, outer drill and pin.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA 02048
MDR Report Key10775880
MDR Text Key214344431
Report Number3014334038-2020-00086
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ76D55
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2020
Date Manufacturer Received12/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-