• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/07/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The ft4 results from all analyzers are around the upper level of the respective normal reference ranges for the assays.Calibration and qc at the investigation site was acceptable.Based on the information provided, a general reagent issue can be excluded.Assays from different manufacturers, in this case abbott, can generate different results.This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used.The investigation did not identify a product problem.
 
Event Description
The initial reporter complained of discrepant results for 1 patient sample tested for elecsys ft4 iii (ft4 iii) on a cobas e801 module compared to the accuraseed method.The results were reported outside of the laboratory where the physician requested additional testing.The sample was submitted for investigation where discrepant results were also identified between the customer¿s e801 module, the abbott method and an e801 module used at the investigation site.Refer to attached data for the results.The customer¿s e801 module serial number was (b)(4).The e801 module used at the investigation site was (b)(4).The ft4 iii reagent lot number used at the investigation site was 460793 with an expiration date of feb-2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10778360
MDR Text Key214390353
Report Number1823260-2020-02753
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2020
Initial Date FDA Received11/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age26 YR
-
-