Model Number 71364009
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 10/06/2020
Event Type  malfunction  
Event Description
It was reported that during an office inspection it was found that the tip of the femoral head impactor missing.No case involved.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The plastic tip of the device is missing, rendering the device inoperable.The device was manufactured in 2019 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD IMPACTOR
Type of Device
PROST, KNEE, PATELLO/FEM, SEMI-CONS, UNC, POROUS, COATED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10779077
MDR Text Key214360561
Report Number1020279-2020-06067
Device Sequence Number1
Product Code MBH
UDI-Device Identifier03596010199256
UDI-Public03596010199256
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364009
Device Catalogue Number71364009
Device Lot Number19EM19292
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/06/2020
Initial Date FDA Received11/03/2020
Supplement Dates Manufacturer Received11/11/2020
Supplement Dates FDA Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-