• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0302
Device Problem Deformation Due to Compressive Stress (2889)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/12/2020
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report deformation to the soft tip.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The steerable guide catheter was prepped and inserted; however, during advancement, the soft tip of the sgc became damaged and was no longer circular.The sgc was removed and it was noted that no damage occurred to the anatomy.A new sgc was used and two clips were successfully implanted, reducing mr to a grade of 2.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other complaints reported from this lot.All available information was investigated and a cause for the deformation due to compressive stress associated with the soft tip deformation cannot be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10779115
MDR Text Key214590916
Report Number2024168-2020-09166
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/09/2021
Device Catalogue NumberSGC0302
Device Lot Number00609U135
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received11/03/2020
Supplement Dates Manufacturer Received01/12/2021
Supplement Dates FDA Received01/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-