(b)(6).Device returned to manufacturer: the product was returned to boston scientific for analysis.Returned product consisted of an ffr comet pressure wire in the shipping tube.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The wire showed that it was separated at 48cm from the tip.The wire shafts showed multiple small bends and kinks throughout the returned wires.With the wire returned separated signal verification could not take place.Upon connecting the device to the analysis support test bench equipment, it was noticed that the coefficient values were not present.This is an indication of the rfid tag not being programmed per document.With no coefficient values programmed the equipment would not recognize the device; therefore, no testing could be completed.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.Device analysis determined the condition of the returned device was consistent with the reported event of a separation.
|