• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION Back to Search Results
Model Number PE074F5
Device Problem Pacing Problem (1439)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/25/2020
Event Type  malfunction  
Manufacturer Narrative
One bipolar pacing catheter without any attached components was returned for evaluation.Continuity testing confirmed a full open condition in the proximal circuit.The distal circuit was found to be continuous.A cut down of the catheter body was performed just proximal of the proximal electrode to expose the pacing lead wires.The proximal lead wire was found to be broken at 1.1 cm from the catheter tip and insulation was not present at the broken section.It was found that the proximal circuit was continuous from the broken lead wire to the proximal connector pin.The balloon inflated clear and concentric with 1.3 cc air and remained inflated for 5 minutes without leakage.The distal windings were found to be partially unraveled.No visible damage from the balloon, proximal windings or catheter body was observed.The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Customer report of pacing issue was confirmed during the analysis.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the swan ganz catheter was unable to pace from the beginning of use after catheter insertion.The catheter was replaced and the problem was solved.It is unknown if the patient had cardiac conduction defect.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR SVC INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500249
MDR Report Key10781987
MDR Text Key214546381
Report Number2015691-2020-14237
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPE074F5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-