Model Number 9733498NAV
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/27/2020
Event Type  Injury  
Manufacturer Narrative
No system checkout will be performed since it was an issue with an instrument.The instrument was discarded by the site.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a spinal procedure to address stenosis.It was reported that a cannula broke and fragmented into right l4.Additional surgery was performed to remove the posterior portion of the cannula.There was no patient harm and the procedure delay was unknown.
 
Manufacturer Narrative
Event date: additional information medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was provided stating that the procedure was delayed by 20-30 minutes.The likely cause of the cannula breakage was the sclerotic nature of the patient's bone along with the surgeon trying to lateralize the trajectory after it was placed in the vertebra.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIGATED PEDICLE ACCESS PACK
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key10785697
MDR Text Key214582446
Report Number1723170-2020-02878
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00643169345171
UDI-Public00643169345171
Combination Product (y/n)N
PMA/PMN Number
K131425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/18/2022
Device Model Number9733498NAV
Device Catalogue Number9733498NAV
Device Lot Number200318F
Was Device Available for Evaluation? No
Date Manufacturer Received11/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient Weight88
-
-