Model Number 4FC12
Device Problem Gas/Air Leak (2946)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/13/2020
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the 4fc12 sheath with lot 0010066558 was returned and analyzed.Visual inspection of the sheath showed the shaft was intact with no apparent issues.Performance testing with the leak tester revealed the pressure decay was in an acceptable range.The shaft and valve were leak tight with no apparent issues.Several flushes and aspirations were performed without issues while the balloon catheter was inserted.There was no blockage along the shaft and the shaft and side port were leak tight.In conclusion, the reported air ingress issue was not confirmed through testing.The sheath passed the returned product inspection as per specification.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, air bubbles were noted after the balloon catheter was inserted in the sheath.The balloon catheter was re-inserted into the sheath without resolving the issue.The sheath was then replaced which resolved the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10786164
MDR Text Key214585808
Report Number3002648230-2020-00572
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00763000217136
UDI-Public00763000217136
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/07/2022
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0010066558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2020
Initial Date FDA Received11/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age39 YR
-
-