SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOME AC POWER SUPPLY; EXTERNAL POWER SUPPLY Back to Search Results
Catalog Number 295600-001
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2020
Event Type  malfunction  
Manufacturer Narrative
The freedom home ac power supply will be evaluated by syncardia.The results will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom home ac power supply had a damaged cord.
 
Manufacturer Narrative
Customer reported that freedom home ac power supply cord was separated from the power supply brick with no comment on functionality.During visual inspection, damage to the power cord with exposed wires was confirmed.Additionally, one of the four power supply rubber feet was found to be missing.The investigation confirmed the customer-reported issue.The observed damage did not affect the functionality of the power supply and was mostly likely caused by rough handling.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOME AC POWER SUPPLY
Type of Device
EXTERNAL POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10786715
MDR Text Key214778280
Report Number3003761017-2020-00222
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295600-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
-
-