SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2020
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The customer, a syncardia authorized distributor, reported that the companion 2 driver exhibited a system malfunction alarm while supporting a patient.The patient was switched to a backup driver and there was no reported adverse patient impact.
 
Manufacturer Narrative
Alarm data history review found a system malfunction alarm recorded in the driver's alarm history confirming the customer reported issue.Visual inspection of internal and external components found no evidence of damage or abnormalities.The driver passed incoming functional testing.To replicate the customer reported system malfunction, the driver was power cycled and approximately one minute after start up the driver produced a system malfunction notification and alarm.This was repeated four more times, replicating the customer reported issue.The root cause was determined to be the pressure sensor board.The root cause of the customer reported system malfunction alarm is the pressure sensor board.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.Ce 5393 (b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10786744
MDR Text Key214770122
Report Number3003761017-2020-00221
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 10/19/2020
Initial Date FDA Received11/04/2020
Supplement Dates Manufacturer Received10/19/2020
Supplement Dates FDA Received02/24/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-