JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Vitrectomy (2643); No Code Available (3191)
Event Date 09/24/2020
Event Type  Injury  
Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced in the initial surgery.If explanted, give date: not applicable, as lens was removed/replaced in the initial surgery.(b)(4).Device evaluation: the complaint sample was not returned to the manufacturing site (the device was discarded); therefore, product testing could not be performed, and the customer's reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.There were no discrepancies found during the mrr (manufacturing record review).The search revealed one complaint from this production order number, however the reported issue is unrelated this reported complaint and no product deficiency was identified in that case.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that after a model pcb00 intraocular lens was fully inserted into the patient's eye, vitreous was discovered.The lens was removed in the same procedure, and a vitrectomy was required.No further information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
Manufacturer (Section G)
JOHNSON & JOHNSON SURGICAL VISION, INC.
road 402 north, km 4.2
anasco industrial park
anasco PR 00610
Manufacturer Contact
somyata nagpal
1700 east st. andrew place
santa ana, CA 92705
7142478200
MDR Report Key10788550
MDR Text Key214758120
Report Number2648035-2020-00814
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474558298
UDI-Public(01)05050474558298(17)230810
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P980040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Type of Report Initial
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/10/2023
Device Model NumberPCB00
Device Catalogue NumberPCB0000215
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/08/2020
Initial Date FDA Received11/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-