ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71992-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Unspecified Infection (1930); Skin Inflammation (2443)
Event Date 07/28/2020
Event Type  Injury  
Manufacturer Narrative
At this time product has not yet been returned.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor products.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A skin reaction was reported with the use of the adc freestyle libre 2 sensor.The customer reported symptoms of infection and inflammation at the sensor site and had contact with a healthcare provider.The customer received cefuorxime 500 mg for treatment.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Visual inspection has been performed on the returned sensor patch (serial number (b)(4)) and no physical damage was observed.No issues were observed with the returned adhesive.Sharp was not returned.No malfunction or product deficiency was identified, therefore, the issue is not confirmed.
 
Event Description
A skin reaction was reported with the use of the adc freestyle libre 2 sensor.The customer reported symptoms of infection and inflammation at the sensor site and had contact with a healthcare provider.The customer received cefuorxime 500 mg for treatment.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key10788826
MDR Text Key214730402
Report Number2954323-2020-09827
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/30/2021
Device Model Number71992-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2020
Initial Date Manufacturer Received 08/13/2020
Initial Date FDA Received11/04/2020
Supplement Dates Manufacturer Received12/09/2020
Supplement Dates FDA Received12/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age19 YR
Patient Weight54
-
-