(b)(4); batch # unk.Date of event is 2020.Event day and event month were not reported.A manufacturing record evaluation was performed for the finished device lot number and no non-conformances were identified.Attempts are being made to retrieve the device.To date, the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.
|
(b)(4).Date sent: 1/6/2021 d4: batch # u93v2p b3: component code: mechanical (g04) investigation summary the analysis results found that the el5ml device was received with one jaw disengaged from the cam.This condition would not allow the jaws to collapse in order to form the clips.No functional test was performed due the condition of the device.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.In addition, four clips were found inside the clip track.No conclusion could be reached as on what caused the jaw to become disengaged.Although there is no direct evidence of use error, the instructions for use do contain the following: "caution: possible causes for the condition of the jaw disengaged from the cam may be inadvertent force, twisting or pressure being placed on the device jaws, using the jaws of the device as a dissector/retractor or damage to the jaws while entering the trocar." it should be noted that as part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|