Device evaluated by mfr: visual inspection revealed that the gauge needle was at 0 atm when received.No visual defects were encountered in the device.Functional inspection revealed that a functional test of the complaint device was carried out using a bsc stopcock; the purpose of the analysis is to ensure that the unit meets the required specification.The flexcil line was easily attached to the stopcock.The unit was pressurized, however, the gauge was not reading, the needle remained in the position of 0 atm.No leaks nor damages in the device were noted.
|
Reportable based on device analysis completed on (b)(6) 2020.It was reported that the device kept on slipping off the delivery system.During angigram procedure, an encore 26 inflation device was selected for use and was screwed onto the stent delivery system.However, it kept on slipping off the delivery system and unable the action to be pulled to negative.The physician tried to do the same process couple of times but device kept on slipping off.It was replaced with another of the same device and used onto the same stent, which worked perfectly.There were no complications reported and the patient is in normal condition after the procedure.However, device investigation revealed that the gauge was not reading and the needle remained in the position of 0 atm though the unit was pressurized.
|